Section 1 Specifications - Hach DR 5000 User Manual

Spectrophotometer
Hide thumbs Also See for DR 5000:
Table of Contents

Advertisement

Section 1
Performance specifications
Operating mode
Source lamp
Wavelength range
Wavelength accuracy
Wavelength reproducibility
Wavelength resolution
Wavelength calibration
Wavelength selection
Scanning speed
Spectral bandwidth
Photometric range
Photometric accuracy
Photometric linearity
Stray light
Data storage
User programs
Physical and environmental specifications
Width
Height
Depth
Weight
Operating requirements
Storage requirements
Additional technical data
Mains connection
Interfaces
Water-resistant to
Enclosure
Safety class
Specifications
Specifications are subject to change without notice.
Transmittance (%), Absorbance and Concentration
Gas-filled Tungsten (visible) and Deuterium (UV)
190–1100 nm
± 1 nm in Wavelength Range 200–900 nm
< 0.1 nm
0.1 nm
Automatic
Automatic, based on method selection
900 nm/min 1 nm steps
2 nm
± 3.0 Abs in Wavelength Range 200–900 nm
5 mAbs at 0.0–0.5 Abs
1% at 0.50–2.0 Abs
< 0.5%–2 Abs
< = 1% at > 2 Abs
with neutral glass at 546 nm
KI-solution at 220 nm
> 3.3 Abs / < 0.05%
1000 measured values (result, date, time, sample ID, user ID)
50
450 mm (17.7 in)
200 mm (7.9 in)
500 mm (19.7 in)
15.5 kg (34.2 Ibs)
10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)
–25–60 °C (–13–140 °F) max. 80% relative humidity (non-condensing)
100–240 V~; 50/60 Hz; automatic changeover
Use only shielded cable with max. length of 3 m.
1 x USB type B
2 x USB type A
IP 31
Safety class I
7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents