Section 4

4.1 Power the instrument on and off

4.2 Language selection

4.3 Self-Check

Start Up
1. Plug in power supply.
2. Close the empty cell compartment.
3. Turn the instrument on by pressing the power switch on the
Note: Do not turn the instrument off and on in rapid succession. Always
wait about 20 seconds before turning the instrument on again, otherwise
the electronic and mechanical systems will be damaged.
The spectrophotometer software includes several language
options. The first time the instrument is turned on, the language
selection screen will appear.
1. Select the desired language.
2. Press OK to confirm the language selection. The self check will
Changing the language setting
The instrument functions in the selected language until the option is
changed.
1. While turning the instrument on, touch the screen at any point
2. Select the required language.
3. Press OK to confirm. The test program subsequently starts
Each time the instrument is powered up, a series of diagnostic tests
are performed automatically to ensure operation of major system
components.
This procedure, which takes approximately two minutes, checks the
system, lamp, filter adjustment, wavelength calibration and voltage.
Each test which functions correctly is confirmed with a check mark.
The Main Menu is displayed when power up diagnostics are
completed.
Note: Further error messages during self check, see
127.