Advanced Operations

6.1.1.1 Single wavelength settings

If the Single Wavelength mode is selected, the following
parameters can be defined:
Units:
Select the required unit from the list and press Next.
Note: Units of measure not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then New.
Wavelength (single wavelength program type):
Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from
190–1100 nm.
Press Next to proceed.
Concentration resolution (number of decimal places)
Select the required number of decimal places from the displayed
list and press Next.