Advanced Operations
4. Press Return to return to the scan mode.
Note: The Integral is shown instead of the date on the display.
Note: For the next scan measurement the setting for the Integral will be
On.
Scale & units
1. Press Scale & Units.
2. Highlight the required units (Abs or %T).
3. Highlight Auto or Manual scaling on the graph´s y-axis.
Note: If manual scaling is selected, use the alphanumeric keypad to set the
limits y
and y
. The graph is adjusted to display only the values in the
min.
max.
selected range. If automatic scaling is selected, the instrument sets the
limits automatically so that the total range can be displayed.
4. Press OK.
5. Press Return to return to the scan mode.

6.6.2 Wavelength scan reading

After the scanning parameters have been selected, the baseline
must be scanned. Changing any of the scanning parameters
requires a new baseline scan. When the baseline has been
scanned, the instrument is ready to scan one or more samples.
1. Press Wavelength Scan in the Main Menu.
2. Insert the blank cuvette/cell into the cuvette/sample cell holder.
97