Agilent Technologies 34410A Service Manual page 100

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures
AC Current Flatness Calibration Procedure
Review the
Considerations"
Configuration:
All AC adjustments use the 3 Hz bandwidth measurement filter
1 Configure each function and range shown in the adjustment table
2 Apply the input signal shown in the "Input" column of the table.
Always complete tests in the specified order as shown in the appropriate table.
N O TE
3 Enter the actual applied input current amplitude (see
4 Repeat steps 1 through 3 for each flatness adjustment point shown in
5 Repeat steps 1 through 4 for each input voltage range table 100 A,
6 Store the new calibration constants
7 Verify the AC Current Flatness adjustments using the verification
Each current and frequency in the flatness adjustment procedure takes less than 10
N O TE
seconds to complete.
100
"Test Considerations"
on page 93 sections before beginning this procedure.
AC Current
below.
Values"
on page 78).
a Successful completion of each adjustment value is indicated by the
message CAL SUCCEEDED flashing in the display.
b If a problem is encountered, the display will flash the message
CAL FAILED. Check the input value, range, function, and entered
adjustment value to correct the problem and repeat the adjustment
step.
the table.
1 mA, 10 mA, 100 mA and 1 A.
page 78.
procedures beginning on
on page 63 and
("Storing the Calibration Constants"
page
71.
34410A/11A/L4411A Service Guide
"Flatness Adjustment
"Entering Adjustment
on