Agilent Technologies 34410A Service Manual page 87

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

AC Voltage Gain Calibration Procedure
Review the
Considerations"
Configuration:
1 Configure each function and range shown in the adjustment table
2 Apply the input signal shown in the "Input" column of the table.
Always complete tests in the specified order as shown in the appropriate table.
N O TE
3 Enter the actual applied rms input voltage
4 Repeat steps 1 through 3 for each gain adjustment point shown in the
5 Store the new calibration constants
34410A/11A/L4411A Service Guide
Input
Current
100 A
1 mA
10 mA
100 mA
1 A
2 A
""Test Considerations"
on page 82 sections before beginning this procedure.
AC Voltage
below.
(see
"Entering Adjustment Values"
a Successful completion of each adjustment value is indicated by the
message CAL SUCCEEDED flashing in the display.
b If a problem is encountered, the display will flash the message
CAL FAILED. Check the input value, range, function, and entered
adjustment value to correct the problem and repeat the adjustment
step.
table.
page 78.
Instrument Settings
Function
Range
DC Current
100 A
1 mA
10 mA
100 mA
1 A
3 A
on page 63 and
"Gain Adjustment
on page 78).
("Storing the Calibration Constants"
Calibration Procedures
3
on
87