DC Current Gain Verification Test
1 Make sure you have read
2 Connect the calibrator to the input terminals.
3 Select each function and range in the order shown below. Provide the
4 Make a measurement and return the result. Compare measurement
Input
Current
100 A
DC Current
1 mA
10 mA
100 mA
1 A
2 A
34410A/11A/L4411A Service Guide
a For the 34410A/11A use the front panel input terminals and select
the Front input terminals with the Front/Rear switch.
input shown in the table below.
results to the appropriate test limits shown in the table. (Be certain to
allow for appropriate source settling when using the Fluke 5720A.)
[1]
Function
Range
100 A
1 mA
10 mA
100 mA
1 A
3 A
[1] Select 100 NPLC.
Q: Quick performance verification test points.
"Test Considerations"
Quick
Check
24 hour
± 0.03 µA
± 0.13 µA
Q
± 2.7 µA
± 14 µA
Q
± 560 µA
± 2.6 mA
Calibration Procedures
on page 63
Error from Nominal
90 day
± 0.065 µA
± 0.075 µA
± 0.36 µA
± 0.56 µA
± 5 µA
± 7 µA
± 35 µA
± 55 µA
± 900 µA
± 1.1 mA
± 3 mA
± 3.6 mA
3
1 year
67