Agilent Technologies 34410A Service Manual page 68

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures
Ohms Gain Verification Test
Configuration:
1 Make sure you have read
2 Set the 4-Wire Ohms function.
3 Connect the calibrator to the input terminals.
4 Select each range in the order shown below. Provide the resistance
Input
Resistance
Function
100
4–Wire Ohms
1 k
10 k
100 k
1 M
10 M
100 M
68
4–Wire Ohms (CONFigure:FRESistance)
a For the 34410A/11A use the front panel input terminals and select
the Front input terminals with the Front/Rear switch.
value indicated. Compare measurement results to the appropriate test
limits shown in the table. (Be certain to allow for appropriate source
settling.)
[1]
Range
100
1 k
10 k
100 k
1 M
10 M
100 M
[1] Select 100 NPLC.
[2] Verify only, no adjustment. 2
Q: Quick performance verification test points.
"Test Considerations"
Quick
Check
24 hour
± 6 m
Q
± 25 m
± 250 m
± 2.5
± 30
Q
± 1.1 k
± 201 k
Wire Ohms.
34410A/11A/L4411A Service Guide
on page 63.
Error from Nominal
90 day
± 12 m
± 14 m
± 80 m
± 110 m
± 800 m
± 1.1
± 8
± 11
± 110
± 130
± 3.1 k
± 4.1 k
± 601 k
± 801 k
1 year