Agilent Technologies 34410A Service Manual page 88

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures
6 Verify the AC Voltage Gain adjustments using the verification
Each range in the gain adjustment procedure takes less than 6 seconds to complete.
N O TE
AC Current Gain Calibration Procedure
Review the
Considerations"
Configuration:
1 Configure each function and range shown in the adjustment table
2 Apply the input signal shown in the "Input" column of the table.
Always complete tests in the specified order as shown in the appropriate table.
N O TE
3 Enter the actual applied rms input current
88
procedures beginning on
Vrms
100 mV
1 V
10 V
100 V
210 V
"Test Considerations"
on page 82 sections before beginning this procedure.
AC Current
below.
(see
"Entering Adjustment Values"
a Successful completion of each adjustment value is indicated by the
message CAL SUCCEEDED flashing in the display.
b If a problem is encountered, the display will flash the message
CAL FAILED. Check the input value, range, function, and entered
page
70.
Input
Instrument Settings
Frequency
Function
1 kHz
AC Volts
1 kHz
1 kHz
1 kHz
1 kHz
on page 63 and
on page 78).
Range
100 mV
1 V
10 V
100 V
750 V
"Gain Adjustment
34410A/11A/L4411A Service Guide