AC Voltage Flatness Calibration Procedure
Review the
Considerations"
Configuration:
The 100V AC range is adjusted with 50Vac input. All AC adjustments uses
the 3 Hz bandwidth measurement filter
1 Configure each function and range shown in the adjustment table
2 Apply the input signal shown in the "Input" column of the table.
Always complete tests in the specified order as shown in the appropriate table.
N O TE
3 Enter the actual applied voltage input amplitude (see
4 Repeat steps 1 through 3 for each flatness adjustment point shown in
5 Repeat steps 1 through 4 for each input voltage range table 100 mV,
6 Store the new calibration constants
7 Verify the AC Flatness adjustments using the verification procedures
Each range in the flatness adjustment procedure takes less than 10 seconds to complete.
N O TE
34410A/11A/L4411A Service Guide
"Test Considerations"
on page 93 sections before beginning this procedure.
AC Voltage
below.
Values"
on page 78).
a Successful completion of each adjustment value is indicated by the
message CAL SUCCEEDED flashing in the display.
b If a problem is encountered, the display will flash the message
CAL FAILED. Check the input value, range, function, and entered
adjustment value to correct the problem and repeat the adjustment
step.
the table.
1 V, 10 V, and 100 V.
page 78.
beginning on
page
70.
Calibration Procedures
on page 63 and
"Flatness Adjustment
("Storing the Calibration Constants"
"Entering Adjustment
on
3
95