Agilent Technologies 34410A Service Manual page 70

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures
AC Volts Verification Test
Configuration:
1 Make sure you have read
2 Connect the calibrator to the input terminals.
3 Set the AC Volts function and the 3 Hz input filter. With the slow filter
4 Select each range in the order shown below. Provide the indicated input
Vrms
100 mV
100 mV
100 mV
1 V
1 V
1 V
10 V
10 V
10 V
10 V
100 V
100 V
50 V
750 V
210 V
70 V
Q: Quick performance verification test points.
70
AC Volts (CONFigure[:VOLTage]:AC)
LF 3 HZ:SLOW ([SENSe:]VOLTage:AC:BANDwidth 3)
a For the 34410A/11A use the front panel input terminals and select
the Front input terminals with the Front/Rear switch.
selected, each measurement takes 2.5 seconds to complete.
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
Input
Frequency
1 kHz
100 mV
50 kHz
100 mV
300 kHz
100 mV
1 kHz
1 V
50 kHz
1 V
300 kHz
1 V
10 Hz
10 V
1 kHz
10 V
50 kHz
10 V
300 kHz
10 V
1 kHz
100 V
50 kHz
100 V
300 kHz
100 V
1 kHz
750 V
50 kHz
750 V
300 kHz
750 V
"Test Considerations"
Quick
Range
Check
± 40 µV
Q
± 90 µV
± 1.50 mV
± 400 µV
± 900 µV
± 15 mV
± 4.03 mV
± 4 mV
Q
± 9 mV
± 150 mV
Q
± 40 mV
± 90 mV
± 1 V
± 615 mV
± 405 mV
± 4.45 V
on page 63.
Error from Nominal
24 hour
90 day
± 80 µV
± 140 µV
± 1.70 mV ± 1.70 mV
± 800 µV
± 1.4 mV
± 17 mV
± 8.03 mV ± 9.03 mV
± 8 mV
± 14 mV
± 170 mV
± 80 mV
± 140 mV
± 1.1 V
± 915 mV
± 564 mV
± 4.59 V
34410A/11A/L4411A Service Guide
1 year
± 90 µV
± 150 µV
± 900 µV
± 1.5 mV
± 17 mV
± 9 mV
± 15 mV
± 170 mV
± 90 mV
± 150 mV
± 1.1 V
± 990 mV
± 585 mV
± 4.59 V