Frequency Gain Verification Test
Configuration:
1 Make sure you have read
2 Select the Frequency function, default range, and minimum resolution
3 Connect the Agilent 33220A to the input terminals.
4 Select each range in the order shown below. Provide the input voltage
Input
Vrms
10 mV
7 V
Q: Quick performance verification test points.
34410A/11A/L4411A Service Guide
Frequency (CONFigure:FREQuency DEF, MIN)
(1 second aperture).
a For the 34410A/11A use the front panel input terminals and select
the Front input terminals with the Front/Rear switch.
and frequency indicated. Compare measurement results to the
appropriate test limits shown in the table. (Be certain to allow for
appropriate source settling.)
Frequency
Range
40 Hz
100 mV
10 kHz
10 V
"Test Considerations"
Quick
Check
24 hour
± 20 mHz
Q
± 0.5 Hz
Calibration Procedures
on page 63.
Error from Nominal
90 day
1 year
± 24 mHz
± 28 mHz
± 0.6 Hz
± 0.7 Hz
3
69