1
Specifications
Performance Versus Integration Time – 60Hz (50Hz) Power line frequency
Integration Time
Number of Power
ppm Range
Line Cycles (NPLC)
[6]
0.001
[6]
0.002
0.006
0.02
0.06
0.2
1
2
10
100
[ 1 ] Resolution is defined as the typical 10 VDC range RMS noise.
[ 2 ] Normal mode rejection for power–line frequency ± 0.1%.
[ 3 ] Maximum rate for DCV, DCI, and 2–Wire resistance functions
[ 4 ] Autozero on for => 1 NPLC.
[ 5 ] For power–line frequency ± 1% 75 dB and for ± 3% 55 dB.
[ 6 ] Only for 34411A /L4411A.
Transfer Accuracy (Typical)
All DC volts, <0.12 A DC Current, < 1.2 M : (24 hour % of range error) / 2)
All other DC current and resistance: (24 hour % of range error + % of reading)/2
Conditions:
20
Resolution
NMR db
[1]
30
0
15
0
6
0
3
0
1.5
0
0.7
0
0.3
55
[5]
0.2
110
[5]
0.1
110
[5]
0.03
110
(using zero settling delay, autozero off, etc.).
Basic dc accuracy specifications (previous page) include RMS noise at 100 NPLC.
For <100 NPLC, add appropriate "RMS Noise Adder" to basic accuracy specification.
- Within 10 minutes and ±0.5 °C
- Within ±10% of initial value.
- Following a 2–hour warm–up.
- Fixed range.
- Using 10 NPLC.
- Measurements are made using accepted metrology practices.
[2]
Readings /
[3]
Second
50,000
25,000
10,000
3000
1000
300
60(50)
30(25)
6(5)
0.6(0.5)
RMS Noise Adder % range
DCV
DCV
10, 1000 V
1, 100 V
Resistance
1K, 10K ohm
0.0060
0.0100
0.0060
0.0030
0.0012
0.0040
0.0006
0.0030
0.0003
0.0020
0.0002
0.0015
0.0
0.0001
0.0
0.0001
0.0
0.0
0.0
0.0
34410A/11A/L4411A Service Guide
[4]
DCV
0.1 V
Resistance
100 ohm
DCI
1 amp
0.1000
0.0600
0.0600
0.0300
0.0200
0.0150
0.0010
0.0010
0.0005
0.0