Agilent Technologies 34410A Service Manual page 84

6 1/2 digit multimeter
Hide thumbs Also See for 34410A:
Table of Contents

Advertisement

3
Calibration Procedures
DC Voltage Gain Calibration Procedure
Review the
Considerations"
Configuration:
1 Configure each function and range shown in the adjustment table
2 Apply the input signal shown in the "Input" column of the table.
Always complete tests in the specified order as shown in the appropriate table.
N O TE
3 Enter the actual applied input voltage
4 Repeat steps 1 through 3 for each gain adjustment point shown in the
5 Store the new calibration constants
6 Verify the DC Voltage Gain adjustments using the verification
Each range in the gain adjustment procedure takes less than 2 seconds to complete.
N O TE
84
"Test Considerations"
on page 82 sections before beginning this procedure.
DC Voltage
below.
(see
"Entering Adjustment Values"
a Successful completion of each adjustment value is indicated by the
message CAL SUCCEEDED flashing in the display.
b If a problem is encountered, the display will flash the message
CAL FAILED. Check the input value, range, function, and entered
adjustment value to correct the problem and repeat the adjustment
step.
table.
page 78).
procedures beginning on
on page 63 and
on page 78).
("Storing the Calibration Constants"
page
66.
34410A/11A/L4411A Service Guide
"Gain Adjustment
on