Measuring Frequency And Period - Tektronix DMM4040 User Manual

Table of Contents

Advertisement

2ND MEAS Cycles the secondary display through the measurement functions

Measuring Frequency and Period

The Meter measures the frequency, or period of ac signals between 3 Hz and 1 MHz,
applied between the Meter's HI and LO connectors.
The G key not only activates the Meter's Frequency/Period function, but also toggles
the Meter's primary display between the signal's frequency and period measurement.
Therefore, whether a frequency or period measurement appears after pressing G
depends on the condition this function was left in the last time it was used.
To make a frequency measurement:
1. Press G.
If S is displayed, press G again to switch the primary display to frequency.
2. Connect the Meter to the signal as shown in Figure 4-1.
To make a period measurement:
1. Press G.
If HZ is displayed, press G again to switch the primary display to period.
2. Connect the Meter to the signal as shown in Figure 4-1.
Each press of G switches the measurement between frequency and
period.
The available impedance settings are presented in sets of three
values. To move to a higher set of impedance values, press ++
-->. Press <-- -- to go to a lower set of impedance values.
3. With the impedance highlighted, press the soft key under the
value selected.
listed below, and then off. When a second measurement function is
selected, the 2ND MEAS soft key label is highlighted.
VDC - Displays the dc voltage on which the ac signal may be riding.
Note
ACV/DCV dual mode should not be used at frequencies
below 10 Hz.
Frequency - Displays the frequency of the ac signal applied to the
Input HI and LO connectors of the Meter.
0.000000
F1
F2
F3
Note
Making Measurements

Measuring Frequency and Period

Hz
F4
F5
4
caw06f.eps
4-7