Time Adjustment; Troubleshoot - Topcom BPM ARM 5500 XLD User Manual

Fully automatic, arm mounted blood pressure monitor
Hide thumbs Also See for BPM ARM 5500 XLD:
Table of Contents

Advertisement

TOPCOM BPM Arm 5500 XLD
9.1

Time adjustment

PUL/min

10 Troubleshoot

Error message
on display
7
Er0
Er1
Er2
Er3
Er4
Er5
Er6
Er7
Er8
Er9
ErA
10
To adjust the date an time:
Press the start/stop
simultaneously. The month will blink on the display.
Press the memory button
the month.
Press the start/stop button
The day will blink on the display.
Change the day, hours and minutes as described above,
using the memory button
stop button
After setting the minutes, the device is ready to use.
Reason
Batteries low
Pressure system is unstable before
measurement.
Systolic pressure detect fail
Diastolic pressure detect fail
Pneumatic system blocked or cuff is
too tight during inflation
Pneumatic system leakage or cuff is
too loose during inflation
Inflation above 300mmHg
Over 3 minutes while pressure
above 15 mmHg
EEPROM accessing error
Device parameter check error
MCU self-verity error
Span error: pressure sensor
parameter error
and memory button
9
successively to change
M
8
to confirm the month.
9
to change and the start/
M
8
to confirm every setting.
9
Solution
Change the batteries
Don't move and try again.
Place the cuff correctly and
try again.
Measure again after 5
minutes. If the monitor is still
abnormal, please contact the
local distributor or the
factory.