Troubleshooting
Fieldbus Error Messages
DDP Error 1: Unable to load DD. Press OK to
return to Block List
There has been a loss of communication with one
or more devices on the segment. Press OK to
return to the Live Device List.
The 375 has been disconnected from the segment.
Press RETRY to attempt to reconnect to the
segment. Press CANCEL to return to the FF app
main menu. Press EXIT to exit the FF App.
SM ERROR: FAILED RESPONDER IDENTIFY
FFLIB: Incompatible DD Static Description
FBLIB: Device Upload aborted
FMS OTHER: Parameter Check
FMS OTHER: Exceeds Parameter Limits
FMS OTHER: Wrong Mode For Request
SM ERROR: REMOTE ERROR INVALID STATE
DD Error..."DD file corrupted"...Reinstall Device
revision <manufacture name, device name, dev.
rev. #, and
DD rev. #>...See the Programming Interface for
details.
Description
Using the Easy Upgrade Programming Utility,
check to see if this DD is available to load
into your 375. If not, contact the device
manufacturer.
Check connection to fieldbus devices.
Verify that the 375 lead set is still securely
connected to the FF Segment and tap
RETRY.
There is a problem sustaining
communications with the device.
Contact 375 Field Communicator Technical
Support personnel.
The device description (DD) does not match
the actual device correctly.
Use the Easy Upgrade Programming Utility to
see if a new DD is available for that device.
Contact 375 Field Communicator Technical
Support personnel.
Communications with a device has timed out.
This should normally not happen, unless a
host was selected. Contact 375 Field
Communicator Technical Support personnel.
The parameter being entered is invalid for
this device. Check with the device supplier.
Make sure the value entered is valid for that
specific device parameter.
The parameter being entered is outside the
range of valid values for the device
parameter. Check with the device supplier.
Make sure the value entered is in the correct
range for that device parameter.
The block is in the wrong mode to accept the
parameter change being requested. See
"Change Modes" on page 4-13 for
information on how to change block modes.
An attempt to change the address of the
device serving as the LAS on the segment is
being made. Changing the device address of
the LAS is not recommended.
The checksum of the DD does not match with
that described in the DD header file. This
error may be corrected by connecting the 375
Field Communicator to the Easy Upgrade
Programming Utility and running a Refresh
operation.