Emerson 375 User Manual page 58

Field communicator
Hide thumbs Also See for 375:
Table of Contents

Advertisement

3-18
Polling Address
0
0-15
0-63
63
Custom Range
• Poll Using Tag - Lets you enter and poll by the tag
of the device. You are prompted to enter a tag
name when the HART Application starts.
• Poll Using Long Tag - Lets you enter and poll by
the long tag of the device. You are prompted to
enter a tag name when the HART Application
starts. (Only supported in HART revision 6 or
higher devices.) Double-tap when the tag is
truncated to display the full tag.
• Poll Using Unique ID -Lets you poll using the
device's unique ID. For HART revision 7 devices,
the unique ID is the expanded device type, a
hyphen, and device ID. For HART revision 6 and
earlier devices, the unique ID is the manufacturer
ID, device type, a hyphen, and device ID.
4.Tap Enter.
Refer to your device manual for information about
changing a device polling address.
Use
Poll loops that are not multi-dropped.
Poll HART revision 5 or earlier devices on
multi-dropped loops.
Poll HART revision 6 or 7 devices on
multi-dropped loops.
Poll the default address for HART revision 7
WirelessHART adapters.
Poll a unique or non-consecutive group of
addresses (example: "0, 5-7, 12"). Use a
hyphen to specify consecutive addresses
and a comma to separate numbers or
groups of numbers.
HART Functionality