Emerson 375 User Manual page 94

Field communicator
Hide thumbs Also See for 375:
Table of Contents

Advertisement

5-8
Hart Application Error..."data compression dll is
corrupted"...Reinstall System Files...See User's
Manual for details.
Memory Error..."data compression memory
allocation error"...Restart Unit...See User's Manual
for further details.
DD Error..."failure while opening DD"...Reinstall
Device Type x Device Rev x dd Rev x...See
Programming Interface for details.
DD Error..."failure while uncompressing
DD"...Reinstall Device revision <manufacture
name, device name, dev. rev. #, and DD rev. #>
...See Programming Interface for details.
Communication Error..."unexpected return
value"...See User's Manual for solution.
DD Error..."DD file corrupted"...Reinstall Device
revision <manufacture name, device name, dev.
rev. #, and
DD rev. #>...See the Programming Interface for
details.
Forward Compatibility Blocked...DD Device Type x
Device Rev x dd Rev x cannot be used in forward
compatibility mode...x... Do you wish to proceed in
forward compatibility mode?
DD Not Installed...DD device type x device rev x is
not installed on the System Card...See
Programming Interface for details on DD
updates...Do you wish to proceed in forward
compatibility mode?
Do you want to ignore this message while
communicating with this device?
Required functionality is missing from the
data compression component.
The HART Application is unable to acquire
enough memory for data compression.
The system cannot open the DD.
An error is detected while uncompressing the
DD.
An unexpected value is returned from a
function call.
The checksum of the DD does not match with
that described in the DD header file. This
error may be corrected by connecting the 375
Field Communicator to the Easy Upgrade
Programming Utility and running a Refresh
operation.
The HART Application attempts to use a DD
to run in forward compatibility mode and finds
that the DD is prevented from running in
forward compatibility mode.
The DD for the connected device is not in the
375. If your System Card is licensed for Easy
Upgrade, you can add DDs to your 375 from
the Easy Upgrade Programming Utility.
The HART Application lets you ignore some
response code warning messages, such as
Update Failure, while communicating with a
device. The next time you communicate with
the device, this message will appear.
Troubleshooting