10.3. Initialization Mode

Initialization restores the contents of the FRAM to the default values.
Three types of initialization are available as shown below.
Initialization mode
RAM initialization
General functions
initialization
initialization

10.3.1. RAM initialization / General functions initialization

Step 1 While pressing and holding the ENTER key, press the F key.
fnc is displayed to indicate that the indicator will enter the general function
mode.
To go back to the weighing mode, press the ESC key.
Step 2 While pressing and holding the ZERO key, press the ENTER key.
1Chc is displayed to indicate that the indicator will enter the check mode.
Step 3 Press the ∧ or ∨ key to select the initialization mode ( ini ) and press the
ENTER key.
Step 4 Press the ∧ or ∨ key to select the item to be initialized and press the ENTER
key.
Step 5 All the status LEDs blink to prompt a confirmation. To initialize, press and hold
the ENTER key for three seconds or more.
After initialization, the indicator is reset and all the display segments are ON.
And the indicator enters the weighing mode.
To exit from this mode without performing initialization, press the ESC key.
Display
ini ra
inifnc
All data
iniall
Initializes the contents of the RAM only.
The zero and tare values will be restored
to 0.
Initializes the contents of the general
functions stored in the FRAM.
Initializes all the contents in the FRAM.
The calibration-related data is also initialized.
So, calibration is required before using
the indicator again.
Page 41
Description
AD-4408A