7.3. Calibration-Related Functions (Calf)

Step 1 Press the ON/OFF key to turn the indicator ON (weighing mode) if it is in the
OFF mode.
Remove the CAL key cover and press the CAL key.
Cal is displayed to indicate that the indicator enters the calibration mode.
Step 2 Press the ENTER key to display Cal5et . The indicator enters the mode of
calibration with an actual load. To go back to the weighing mode, press the ESC
key.
Step 3 Press the ∧ or ∨ key to select CalfF and press the ENTER key.
Step 4 Press the ∧ or ∨ key to select a function number to be used and press the
ENTER key. The current setting value is displayed.
Step 5 Change the value as necessary. Two methods to change are available,
depending on the function number as below
Selecting a parameter
Inputting the value
After setting, press the ENTER key. The next function number is displayed.
When the value is not to be changed, press the ESC key to return to the
function number display.
Step 6 Press the ESC key.
calibration-related functions is written into the FRAM.
Press the ESC key again to go back to the weighing mode.
NOTE: The blinking decimal point indicates that the current value is not the weight
value.
If a value exceeding the settable range is entered, err dt is displayed and
the input is canceled.
AD-4408A
Method
Only the available parameter is displayed and blinks.
Press the ∧ or ∨ key to select a parameter.
All the digits are displayed and a digit to be changed
blinks.
Press the < or > key to select a digit and
press the ∧ or ∨ key to change the value.
Cal5et
is displayed and the data of the