Calibration; Items Of Calibration Mode - A&D AD-4405A Instruction Manual

Weighing indicator
Table of Contents

Advertisement

5. Calibration

This indicator, converts an input voltage from a load cell to the "mass" value, and
displays it. Calibration is the adjustment function so that the scale (indicator) can
display the weight correctly.

5.1. Items of Calibration Mode

There are four items in the calibration mode.
How to enter the calibration mode:
In the weighing mode, press the CAL key which is located behind the sealing panel.
Cal in is displayed for 2 seconds and then Cal 0 will appear. Use the MODE key to
select an item and use the ENTER key to execute it.
NOTE: Calibration could be started by simultaneously pressing the
keys, instead of the CAL key. The procedure is prohibited when the indicator
has received a certified approval.
Items Required to be Performed
Cal5et
Sets capacity, resolution, decimal point position and display format,
weighing range and unit. These items should be set first in order for the
indicator to function as a weighing instrument. Set values do not need to
be changed again unless the indicator itself is replaced. For details, refer
to "5.2.1. Configuring a Weighing Instrument".
Cal 0
Calibrates zero and span. This is required after installation, to get accurate
data. For details, refer to "5.2.3. Zero Calibration" and "5.2.4. Span
Calibration".
Optional Items to be Performed as Necessary
lnr 0
Performs digital linearization. Refer to "5.4. Digital Linearization
Function".
g 5et
Compensates for acceleration of gravity.
Compensation Function".
Gravity compensation function: Compensates for weighing error
between the calibration location and another weighing location using
gravity acceleration.
In the calibration mode, the keys have functions as follows:
-
Numerical keys.
0
9
The key to display other items.
MODE
The key to restore the numerical value to the initial setting or to change the
CLEAR
parameters.
+/-
The key to display other parameters.
The key to proceed to the next step without changing set values.
ESC
The key to store new calibration data and proceeds to the next step.
ENTER
AD-4405A Weighing Indicator
Page 11
and
ZERO
Refer to "5.5. Gravity
5. Calibration
TARE

Advertisement

Table of Contents
loading

This manual is also suitable for:

Op-03Op-05Op-07Op-08