Basic Weighing Function; Weighing Mode - A&D AD4329A-DLC Instruction Manual

Table of Contents

Advertisement

7. Basic Weighing Function

7.1. Weighing mode

When the AD4329A-DLC is turned on, all the segment of the display illuminate/extinguish for checking the display.
Then the AD4329A-DLC is in weighing mode and starts weighing.
The contents in the display in weighing mode are as follows.
・ Weighing value
・ Positive overload
・ Negative overload
Error display
errAd n
errlC n
errov n
-------
Description
The n-th digital load cell has a problem in the internal circuit.
The n-th digital load cell doesn't respond.
Check if serial number setting is correct.
The load on the n-th digital load cell exceeds the rated capacity.
Check the actual load.
The load exceeds the power-on zero range.
Press [MODE] to check the current load.
-