Configuration
Function
Prompt
Lower Display
Transmitter
XMITTER2
Characterization
ANALYTIC
Analytic Selections
XMITTER2 must be set to
ANALYT for this prompt to
appear
Input 2 High Range Value
IN2 HI
(Linear Inputs only)
Input 2 Low Range Value
IN2 LO
(Linear Inputs only)
Input 2 Ratio
RATIO 2
Input 2 Bias
BIAS IN2
Input 2 Filter
FILTER 2
BURNOUT2
Burnout Protection
Emissivity
EMISSIV2
38
Function
Name
UDC3300 Expanded Model: DC330E User Manual
Selections or
Range of Setting
Upper Display
B TC
S TC
E TC H
T TC H
E TC L
T TC L
J TC H
W TC H
J TC L
W TC L
K TC H
100 PT
K TC L
100 LO
NNM H
200 PT
NNM L
500 PT
NM90 H
RAD RH
NM90 L
RAD RI
NIC TC
LINEAR
R TC
SQROOT
ANALYT
NONE
PH
ORP Mv
(Millivolts)
CONDms (Milli Siemens)
CONDus (Micro Siemens)
RSTVM^ (Megaohms)
TDS PPm (Parts per Million)
TDS PPb (Parts per Billion)
CONCPt (Parts per Thousand)
DO PPm (Parts per Million)
DO PPb
(Parts per Billion)
–999. to 9999. floating
(in engineering units)
–999. to 9999. floating
(in engineering units)
–20.00 to 20.00
–999. to 9999.
(in engineering units)
0 to 120 seconds
NONE DOWN
UP
NO_FS
0.01 to 1.00
Factory
Setting
LINEAR
NONE
1000
0
1.000
0
1
NONE
0.00
5/00