Honeywell TMS 9250 User Manual page 56

Torque measurement system
Table of Contents

Advertisement

6. Appendices
008-0738-00
Step 5a:
Connect Wires
To Connectors
ATTENTION:
ATTENTION:
56
© 2012 – Honeywell International Inc. All rights reserved
Loosen the screws of the connectors on the PCB and connect the respective wires
to the connectors by referring to the below table.
Connector Details
Power
J1
Supply
By default, the power supply cable is connected.
Connector Details
J2
RS232
Connector Details
Analog Voltage /
J3
Current
The output depends on OPTYPE value configured using TMS Toolkit for TMS 9250.
TMS 9250 Product User Manual
Pin Details
Description
+12V
+12VDC
GND
0VDC
EARTH
Shield/Braid
Pin Details
Description
V+
No connection
TXD
RXD
GND
Earth
Pin Details
Description
Voltage / Current
Vout
Output
Iout
No Connection
GND
Signal Ground (0V)
EARTH
Shield/Braid
2
3
5
0
Rev B