Vertical Controls - Tenma 72-6802 User Manual

30mhz dual trace oscilloscope
Table of Contents

Advertisement

Vertical Controls

1 VOLTS/DIV
Controls the CH1/CH2 vertical scale from 
Knob
1mV/DIV to 5V/DIV in 12 steps. 
2 Vertical
Controls the vertical position of traces and light 
POSITION
spots for CH1/CH2. 
Knob
3 ALT/CHOP
Selects how the CH1 and CH2 signals appear in the 
Switch
display, in the DUAL trace mode. 
CHOP
The two signals are chopped and 
displayed simultaneously (generally 
used in slow sweep, 1ms/DIV or 
slower). 
ALT
The two signals are displayed 
alternatively (generally used in fast 
sweep, 0.5ms/DIV or faster). 
4 Vertical
Selects CH1 and CH2 display modes.  
MODE
Switch
CH1/CH2  The CH1 or CH2 signal is displayed 
independently. 
PANEL OVERVIEW
5 VAR Knob
6 CH2 INV
17
18
DUAL 
The CH1 and CH2 signals are displayed 
simultaneously. 
ADD 
The CH1 and CH2 signals are added or 
subtracted, and then the result is 
displayed. For CH1/CH2 
addition/subtraction details, see page32. 
Adjusts the vertical scale. 
At the minimum position, the vertical scale 
becomes 2.5 times wider than the original value 
selected by the VOLTS/DIV knob. For example, if 
the original scale is 1mV/DIV, the adjusted scale 
becomes 2.5mV/DIV. 
At the maximum (CAL) position, there is no 
change in the vertical scale. 
Inverts the CH2 input signal vertically. When the 
Switch
vertical mode switch is in the ADD position, CH2 is 
subtracted from CH1 instead of being added to 
CH1. For CH1/CH2 addition/subtraction details, 
see page32. 