• If oxygen is used with the device, the oxygen flow must be
turned off when the device is not in use.
Explanation of the Warning: When the device is not in operation
and the oxygen flow is left on, oxygen delivered into the
ventilator tubing may accumulate within the device's enclosure.
Oxygen accumulated in the ventilator enclosure will create a risk
of fire.
• If you are using oxygen, the BiPAP S/T must be equipped with
the Respironics Pressure Valve (Part number 302418). Failure to
use the Pressure Valve could result in a fire hazard.
• Oxygen supports combustion. Oxygen should not be used while
smoking or in the presence of an open flame.
• Do not use the device in the presence of a flammable anaesthetic
mixture in combination with oxygen or air, or in the presence of
nitrous oxide.
• Do not use the device if the room temperature is above 95º F
(35°C). If the device is used at room temperatures above 95° F,
the temperature of the airflow may exceed 106º F (41°C), which
could cause irritation to your airway.
• Do not operate the device in direct sunlight or near a heating
appliance because these conditions can increase the temperature
of the air coming out of the device.
• When using this product, IEC 60601-1-1 requirements (safety
requirements for medical electrical systems) must be met.
• For proper use, the power supply must be placed feet down, in
the upright position.
• When the BiPAP S/T is used with a humidifier, position the
humidifier so that the water level in the humidifier is lower than
you, and the humidifier is on the same level or lower than the
BiPAP S/T.
• Do not attempt to wear your mask without the device turned on.
Doing so could result in CO
• If you notice any unexplained changes in the performance of
the BiPAP S/T, if it is making unusual or harsh sounds, if it and/
or the power supply has been dropped or mishandled, if the
enclosure is broken, or if water has entered the unit, discontinue
use and contact your home care provider.
• Repairs and adjustments must be performed by Respironics -
authorized service personnel only. Unauthorized service could
cause injury, invalidate the warranty, or result in costly damage.
• Periodically inspect electrical cords, cables, and the power
supply device for damage or signs of wear.
rebreathing.