Weighing mode
Method 1
Touch & hold
Calibration mode
Power OFF

10.2. Gravity acceleration correction

When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be
calibrated using a calibration weight at the place where it is to be used.
1. Refer to "10.1. Calibration mode" to enter the calibration
mode. The gravity acceleration value is displayed.
2. Tap the
acceleration value setting mode.
3. Change the displayed value using the following switches.
UNITS
SAMPLE
4. Tap the
TARE
returns to the newly stored gravity acceleration value.
5. When calibration using a calibration weight is to be
performed, go to step 3 of "10.3. Calibration using a
weight".
To finish the setting procedure, touch the
or the CAL switch. The scale returns to the weighing mode.
Method 2
Press
Gravity acceleration
correction
Calibration using
a weight
Restoring the factory
set values
or
switch to enter the gravity
TARE
To shift the blinking digit to the right.
To increase the value of the blinking
digit by one.
switch. The display shows "
Tap the
to select an item
and
Tap the
to execute it.
To exit from the calibration mode, touch
the
switch or Press the
switch. The scale turns the power OFF.
end
" and
switch
17
switch
UNITS
switch
TARE
CAL
Tap
Tap
Tap

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sj-30kwpSj-6000wpSj-15kwpIp67

Table of Contents