9-2. Gravity Acceleration Correction

When the balance is first used or has been moved to a different place, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the balance. Change the gravity acceleration value of the balance to the
value of the area where the balance will be used. Refer to the gravity acceleration map
appended to the end of this manual.
 Note
Gravity acceleration correction is not required when the balance is
calibrated using a calibration weight at the place where the balance
is to be used.
1. Press and hold the calibration (CAL) switch until
Cal
appears, and release the switch.
2. The balance displays Cal 0 .
3. Press the TARE key.
The balance displays the gravity acceleration
value stored in the balance.
Use the following keys to change the value.
SAMPLE key
TARE key
ENTER key
UNITS key
4. Press the ENTER key to store the value.
Cal 0
appears.
5. If it is necessary to calibrate the balance using a
calibration weight, proceed to step 3 of 8-1.
6. To finish the setting, press the MODE key.
The balance returns to the weighing mode.
To select the digit to change
the value (blinking).
To change the value of the
blinking digit.
To store the value and
proceed to step 4.
To cancel the operation and
return to step 2.
20
Press and hold
the CAL switch.
Cal
Release the CAL switch.
Cal 0
9. 7 985
Set the value using
the relevant keys.
To store
Cal 0
Returns to
the weighing mode

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ek-3000epEk-12kepEk ep series