If one-point is chosen:
Use 1 [DOWN] and 2[UP] to choose the
a.
CAL gas concentration.
Press 3[ENTER] to enter the concentration.
b.
Apply the CAL gas and press 2[OK], press
c.
3 [SKIP] to exit calibration mode. The Meter
will now calibrate the CO
press 3 [Cancel] to exit calibration mode If
calibration fails, repeat the procedure. If it fails a
second time, have the Meter serviced.
d.
The Meter finishes the procedure then prompts
the user to enter the time until the next
calibration. Enter the amount using 1[DOWN]
and 2[UP] to change the number and
3[ENTER] to store the number.
22
sensor for 2 minutes.
2
If two-point is chosen:
Apply the neutral gas (Nitrogen) or press 3
a.
[Cancel] to exit calibration mode.
b.
When zeroing is complete, remove neutral gas
(Nitrogen) and press 2[OK] or 3 [Cancel]
to exit calibration mode.
c. Apply the CAL gas. Enter the amount using
1[DOWN] and 2[UP] to change the number
and 3[ENTER] to store the number
d. Apply cal gas and press 2[OK] or 3 [Cancel]
to exit calibration mode.
e. The Meter finishes the procedure then prompts
the user to enter the time until the next calibration.
Enter the amount using 1[DOWN] and
2[UP] to change the number and 3[ENTER]
to store the number.
CO
sensor calibration is now complete and the Meter
2
exits calibration mode.
.