Only to older units with firmware up to lss 2.x (324 pages)
Summary of Contents for Fluke 975CK
Page 1
The Meter’s CO and CO sensors can be calibrated by the user or returned to Fluke for service. See “Contacting Fluke”. The recommended calibration interval is 1 month for CO and 1 year for CO...
Page 2
Calibrating the CO and CO Sensors Calibrate the CO and CO sensors together, or separately. To simultaneously calibrate both sensors, If not already in calibration mode, press L, R, and S simultaneously for 3 seconds to enter calibration mode. Press 1 to initiate the calibration procedure.
Page 3
The Meter prompts to apply the CAL Gas. Attach the mixed CO and CO calibration gas canister to the Meter via the supplied hose or press 3 [SKIP] to exit. The Meter calibrates the CO sensor. Press 3 to cancel. When finished, remove gas and press 3 [OK].
Page 4
Regulators are warranted for 2 years from the date of purchase against defects in the materials and assembly, and performance within specification. If the regulator is considered by Fluke to be defective then Fluke will replace the regulator at no charge. If the gas canister leaks or the gas becomes unstable within 3 years from the manufacturing date on the canister, Fluke will replace the canister at no charge.