Ikegami MLW-2420C Instruction Manual page 23

Wide screen lcd type, high quality full hd lcd monitor for medical applications
Table of Contents

Advertisement

G. Offset:
The dark zone of G (green) is adjusted. The black level of G gets lower to-
ward the + side and higher toward the - side.
B. Offset:
The dark zone of B (blue) is adjusted. The black level of B gets lower toward
the + side and higher toward the - side.
RGB. Offset:
The dark zones of all R, G and B are adjusted. The black level of R/G/B gets
lower toward the + side and higher toward the - side.
R. Gain:
The highlight of R (red) is adjusted. The white level of R gets higher toward
the + side and lower toward the - side.
G. Gain:
The highlight of G (green) is adjusted. The white level of G gets higher
toward the + side and lower toward the - side.
B. Gain:
The highlight of B (blue) is adjusted. The white level of B gets higher to-
ward the + side and lower toward the - side.
RGB. Gain:
The highlights of all R, G and B are adjusted. The white level of R/G/B gets
higher toward the + side and lower toward the - side.
Gamma Control:
The gamma level can be adjusted in 0.1 increments in the range of 1.6-2.8.
D14 means a gamma level of DICOM part 14.
User:
A user-set gamma level can be selected. (
signals.)
SAVE:
Select SAVE to put all the above settings in memory.
If the SAVE is not selected, all the previous settings remain.
is a common setting for all the