Barco E-2320 C Installation & User Manual page 51

Hide thumbs Also See for E-2320 C:
Table of Contents

Advertisement

Operation
2. If you have multiple Nio displays of the same type connected
(multi-head system), consider if you wish to equalize the
display you calibrate to a reference display.
The first calibrated display in the system is the reference
display.
Equalization means the luminance of the display you are
calibrating will be matched to the luminance of the reference
display.
If you wish to equalize the display, you must close the
Display Settings dialog and make sure the Equalization
option in the Application Settings dialog is set. See
"Calibration tab" in the description of the Application
Settings dialog.
To calibrate the display using the Barco LCD sensor:
1. If you have multiple displays connected (multi-head system),
select the display you wish to calibrate by clicking on the
corresponding numbered icon in the Display Settings dialog.
2. Click the Calibrate button.
3. A message appears, showing the different calibration
options.
4. Select "Using the BARCO LCD sensor" and click OK.
This option is grayed in case NioWatch does not find a
connected Barco LCD sensor.
E-2320 C
Figure 21: Calibration choices
51