[SOURce:]FREQuency[1]

:MODE

Parameters
Comments
Example
Chapter 8
[SOURce:]FREQuency[1]:MODE <
control the frequency subsystem.
Parameter
Parameter
Name
mode
<
>
discrete
The
mode
parameter has the following meanings:
<
>
– CW
or
FIXed:
Selects single-frequency mode.
selects the sample rate or waveform frequency.
:FIXed]
equivalent.
– FSKey:
Selects frequency shift keying mode.
and the front panel's " Stop Trig/FSK/Gate In" BNC select the two sample rate
or waveform frequencies.
– LIST:
Selects frequency list mode.
sample rate or waveform frequencies.
– SWEep:
Selects frequency sweep mode.
,
and
SPAN
STARt
STOP
range. The
[SOURce:]SWEep
control the sweep.
Executable when Initiated: Query form only
Coupling Group: Frequency
Related Commands:
TRIGger[:STARt]:SOURce
subsystem,
[SOURce:]LIST2
*RST Condition:
SOURce:FREQuency1:MODE FIXed
Setting the Frequency Sweep Mode
FREQ:MODE LIST
mode
>
determines which set of commands
Range of
Type
Values
CW | FIXed | FSKey | LIST |
SWEep
[SOURce:]FREQuency[1]:FSKey
[SOURce:]LIST2:FREQuency
[SOURce:]FREQuency[1]:CENTer
commands set the sample or waveform frequency
,
ARM:SWEep
, and
subsystem,
[SOURce:]SWEep
Sets the frequency sweep mode.
Default
Units
none
[SOURce:]FREQuency[1][:CW |
and
are
CW
FIXed
sets the
TRIGger:SWEep
subsystems
,
[SOURce:]FREQuency[1]
subsystem
Command Reference
,