*STB?

Comments

*TRG

Comments

*TST?

Comments
426
Command Reference
*STB?
returns the value of the Status Byte Register. Bit 6 (decimal weight 64) is
set if a service request is pending.
Register if a service request is generated by a message available (MAV) condition.
*STB?
is a query. Thus, sending the command in response to a MAV condition
will generate Error -410 "Query interrupted".
Executable when Initiated: Yes
Coupling Group: None
Related Commands:
*RST Condition: None
is the command equivalent of the GPIB Group Execute Trigger and the
*TRG
VXIbus Trigger word serial command and has exactly the same effect.
Executable when Initiated: Yes
Coupling Group: None
Related Commands:
*RST Condition: None
*TST?
causes the Agilent E1445A to execute its internal self-test and return a value
indicating the results of the test.
A zero (0) response indicates that the self-test passed. A one (1) response indicates
that the test failed. The failure also generates an error message with additional
information on why the test failed.
When the test completes, all waveform segment and segment sequence definitions
are deleted, and all other commands are set to their
Executable when Initiated: No
Coupling Group: None
*RST Condition: None
STB?
should not be used to read the Status Byte
*SRE
ARM
and
TRIGger
subsystem,
[SOURce:]
commands
values.