State; Unit[:Angle] - Agilent Technologies E1445A User's Manual And Scpi Programming Manual

Arbitrary function generator module scpi programming guide
Hide thumbs Also See for E1445A:
Table of Contents

Advertisement

:STATe

Parameters
Comments
Example

:UNIT[:ANGLe]

Parameters
Comments
Example
Chapter 8
[SOURce:]PM:STATe <
wave output. Phase modulation is always disabled for other waveform shapes.
Parameter
Parameter
Name
<
mode
>
boolean
Executable when Initiated: Yes
Coupling Group: Frequency
Related Commands:
[SOURce:]FUNCtion[:SHAPe
*RST Condition:
SOURce:PM:STATe OFF
Enabling Phase Modulation
FUNC:SHAP SIN
PM:STAT ON
INIT
PM:DEV .78648
[SOURce:]PM:UNIT[:ANGLe] <
[SOURce:]PM[:DEViation]
– DEG:
Degrees
Radians
– RAD:
Parameter
Parameter
Name
units
discrete
<
>
Executable when Initiated: Yes
Coupling Group: None
Related Commands:
[SOURce:]PM[:DEViation]
*RST Condition:
SOURce:PM:UNIT:ANGLe RAD
Setting the Default Angle Units
PM:UNIT:ANGL DEG
mode
>
enables or disables phase modulation for sine
Range of
Type
Values
OFF | 0 | ON | 1
Selects sine wave output.
Enables phase modulation.
Starts output.
Sets deviation to /4.
units
sets the default angle units for subsequent
>
commands. The available default units are:
Range of
Type
Values
DEG | RAD
Sets default units to degrees.
[SOURce:]PM
Default
Units
none
]
Default
Units
none
Command Reference
367

Advertisement

Table of Contents
loading