Photometric analyzer
Table of Contents

Advertisement

2 Installation
2. Calibration with a span filter (this method is available only if you select
a span filter option when you purchase the equipment.
Method One:
1.
Inject zero fluid and set zero as referred in section 3.4 section I
2.
Inject span fluid and set the concentration of the span fluid with
the span procedure referred in section 3.4 section I
Method Two:
1.
Determine the span setting using Method One.
2.
Activate the span filter (as referred in section 3.3.8) section I
3.
Record the display reading (this is the span filter reading and
must be recorded).
4.
You can calibrate the instrument now with the span filter.
Power up the system, and test it as follows:
1. Repeat the Self-Diagnostic Test.
2-4: Part II
Teledyne Analytical Instruments
Model 6020