Maintenance; Fuse Replacement - Teledyne 6020 Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

Part I: Control Unit
Aside from normal cleaning and checking for leaks at the gas connec-
tions, routine maintenance is limited to replacing filter elements and fuses,
and recalibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
4.1
The 6020 requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.
The fuses are located inside the main housing on the Electrical
Connector Panel, as shown in Figure 4-2. To replace a fuse:

Maintenance

MANUAL.

Fuse Replacement

1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse
holder cap, push in, and rotate 1/4 turn. The cap will
pop out a few millimeters. Pull out the fuse cap and
fuse, as shown in Figure 4-1
Teledyne Analytical Instruments
Maintenance 4
Part I: 4-1

Advertisement

Table of Contents
loading

Table of Contents