Part I: Control Unit
Preamp
4.3
All internal components are accessed by unbolting and swinging open
the front cover, as described earlier. The major internal component locations
are shown in Figure 4-2, the cell block is illustrated in Figure 3-2, and the
fuse receptacle is shown in Figure 3-3
The 6020 contains the following major internal components:
•
Customer Interface PCB (Power Supply on bottom surface)
•
Preamp PCB (Contains Microprocessor)
•
Front Panel PCB (Contains Displays)
See the drawings in the Drawings section in back of this manual
for details.
For Optical/Detector Alignments, refer to parts II or III of this manual
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
>3
Call factory for information
Detector
0
OK
1
Failed (open filament, short to ground, no
power.)
2
Unbalance (deterioration of filaments, blocked
tube)
Major Internal Components
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
Teledyne Analytical Instruments
Maintenance 4
Part I: 4-3