Major Internal Components - Teledyne 6020 Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

Part I: Control Unit
Preamp
4.3
All internal components are accessed by unbolting and swinging open
the front cover, as described earlier. The major internal component locations
are shown in Figure 4-2, the cell block is illustrated in Figure 3-2, and the
fuse receptacle is shown in Figure 3-3
The 6020 contains the following major internal components:
Customer Interface PCB (Power Supply on bottom surface)
Preamp PCB (Contains Microprocessor)
Front Panel PCB (Contains Displays)
See the drawings in the Drawings section in back of this manual
for details.
For Optical/Detector Alignments, refer to parts II or III of this manual
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
>3
Call factory for information
Detector
0
OK
1
Failed (open filament, short to ground, no
power.)
2
Unbalance (deterioration of filaments, blocked
tube)

Major Internal Components