Photometric Analyzer
3.
press Enter. The third System screen appears:
4.
The number on the second row is the setpoint of the 4 mA output. It is
analogous to a potentiometer wiper. The number can be set anywhere from 0
to 500. The default is 250, in the middle. At the default setting, the output
should be very close to 4 mA. If not, slowly adjust the number using the Up or
the Down keys until DMM reads 4.00 mA. Enter when done.
5.
should read close to 20 mA. If not, slowly adjust the number using the Up or
Down key until DMM reads 20.0 mA. Enter when done to return to system
menu.
The range of adjustment is approximately +/- 10% of scale (+/- 1.6 ma).
Since the 4 to 20 mA output is tied to the 0 to 1 volt output, this function can be
used to calibrate the 0 to 1 volt output, if the 4 to 20 mA output is not used. By
using a digital Volt meter on the 0-1 Volt output.
3.3.12 Manual control of filter and solenoids
For troubleshooting purposes, you have manual access to control
calibration filter and solenoid on the Analyze mode. To have manual access to
the calibration filter and solenoid:
-Enter the System Menu
-Select MORE on the first and second System menu screens.
-In the last System Menu screen you will see:
ALGORITHM APPLCATION
MODEL OUT_CAL ANLZ
-Select the last field "ANLZ" using the Up/Down switch.
-Press Enter to change the mode of the filter and the solenoid. The se-
quence is as follows:
Using the Right or the Left arrow keys, select MORE and
Select OUTPUT_CAL and Enter
/
4
:
A screen similar to the one above will appear and the DMM
Teledyne Analytical Instruments
_
250
Operation 3
Part I
3-15