Appendix A-1 Specifications - Teledyne 6020 Operating Instructions Manual

Photometric analyzer
Table of Contents

Advertisement

Photometric Analyzer
Appendix
Appendix
A-1 Specifications
6020 Digital Control Module:
Ranges: Three Programmable Ranges, field selectable
within limits (application dependent) and Auto
Ranging
Display: 2 line by 20 alphanumeric VFD accompanied
by 5 digit LED display
Signal Output: Two 0-1V DC (concentration and range ID)
Two 4-20mADC isolated (concentration and
range ID)
RS232
Alarm: Two fully programmable concentration alarm
set points and corresponding Form C, 3 amp
contacts. One system failure alarm contact to
detect power, calibration, zero/span and
sensor failure.
Mounting: Explosion Proof Housing
o
Operating Temperature: 0-50
C
A-1
Teledyne Analytical Instruments