Photometric Analyzer
Fuse Installation:
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to
install the proper fuse as part of installation (See Fuse Replacement in
chapter 4, maintenance.)
Analog Outputs:
the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 2-5.
The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing sample con-
0–1 V dc Range ID:
4–20 mA dc % Range: (-M Option) Current increases linearly with increas-
4–20 mA dc Range ID: (-M Option) 8 mA = Range 1, 12 mA = Range 2,
The fuse holders accept 5 x 20 mm, 4.0 A, T
There are eight DC output signal connectors on
centration, from 0 V at 0% to 1 V at 100%. (Full
scale = 100% programmed range.)
0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3.
ing sample concentration, from 4 mA at 0% to 20
mA at full scale 100%. (Full scale = 100% of
programmed range.)
16 mA = Range 3.
Figure 2-5: Analog Output Connections
Teledyne Analytical Instruments
Part I: Control Unit
Part I: 2-3