The Analyze Function; Signal Output - Teledyne 3010PA Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

Oxygen Analyzer
4.7
When the Analyze function is active, the 3010 is monitoring the
sample gas currently flowing in the Analysis Unit cell block. All undefeated
alarms are ready to activate should their respective setpoints be crossed.
Press the Analyze button to put the analyzer in the Analyze mode.
Normally, all of the functions automatically switch back to the Analyze
function when they have completed their assigned operations. Pressing the
Escape button in many cases also switches the analyzer back to the Ana-
lyze function. Alternatively, you can press the Analyze button at any time
to return to analyzing your sample.
4.8
The standard Model 3010PA Oxygen Analyzer are equipped with two
0-1 V dc analog output terminals accessible on the back panel (one concen-
tration and one range ID). Two isolated 4-20 mA dc current outputs (one
concentration and one range ID), in addition to the voltage outputs, are
optional.
See Rear Panel in chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on range that was defined
as 0–10 % O

The Analyze Function

Signal Output

, then the output would be:
2
Voltage Signal
% O
Output (V dc)
2
0
0.0
1
0.1
2
0.2
3
0.3
4
0.4
5
0.5
6
0.6
7
0.7
8
0.8
9
0.9
10
1.0
Part I: Control Unit
Current Signal
Output (mA dc)
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
Part I: 4-17