Teledyne 3010PA Operating Instructions Manual page 92

Percent oxygen analyzer
Table of Contents

Advertisement

Oxygen Analyzers
A-1 Model 3010PA Specifications
Sample System: Positive pressure service.
Appendix
Packaging: General Purpose Control Unit
• Flush panel mount (Standard).
• Rack mount — Relay rack mounted to
contain either one or two instruments in one
19" relay rack mountable plate (Optional).
Packaging: Explosion Proof Analysis Unit
NEMA 4 Instrument Enclosure. External gas
connector panel with flow indicator (flame
arrestors optional).
Sensor: B-1 Micro-Fuel Cell (standard); others avail-
able.
Cell Block: Nylon (316 stainless steel available).
Ranges: Three user definable ranges 0-1 % to 0-100 %.
Air calibration range 0-25 %.
Autoranging with range ID output.
Vacuum service (optional).
Auto Cal / Auto Zero. Electrically operated
valves.
Alarms: One system-failure alarm-contact to detect
power failure.
Two adjustable concentration threshold alarms
with fully programmable setpoints.
Diagnostics: Start-up or on-demand, comprehensive, self
testing function initiated by keyboard or
remote command.
Appendix
Appendix: A-3