Press Escape to re-enter the Analyze mode using the fixed range.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value)
as established by the operator for that particular range, the
output saturates at 1 VDC (or 20 mA). However, the digital
readout and the RS-232 output continue to read the true
value of the oxygen concentration regardless of the analog
output range.

4.7 The Analyze Function

Normally, all of the functions automatically switch back to the
Analyze function when they have completed their assigned operations.
Pressing the Escape button in many cases also switches the analyzer
back to the Analyze function. Alternatively, you can press the Analyze
button at any time to return to analyzing your sample.

4.8 Signal Output

The standard Model 3000ZA2G Trace Oxygen Analyzer is
equipped with two 0–1 VDC analog output terminals accessible on the
back panel (one concentration and one range ID), and two isolated 4–20
mA DC current outputs (one concentration and one optional range ID).
See Rear Panel in Chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on range that
was defined as 0–100 ppm O
Teledyne Analytical Instruments
L—10
M—1000
H—1000
Mode—FX/MED
or
L—10
M—1000
H—1000
Mode—FX/ HI
, then the output would be:
2
3000ZA2G
50