Oxygen Analyzer
4.1
Once the analyzer has been installed, configure it for your process. To
do this you can:
Before configuration these default values are in effect:
PARAMETER
LO Range
MED Range
HI Range
Auto Ranging
Alarm Relays
Span
Zero
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.

Operation

Introduction

Set system parameters—
Specify a password, if desired, requiring operator to log in.
Establish and start an automatic calibration cycle, if desired.
Calibrate the instrument.
Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.
Set alarm setpoints, and modes (latching, failsafe, etc).
DEFAULT
1 %
5 %
10 %
ON
10 %
(Defeated, HI, Not failsafe, Not latching)
20.9 %
(Auto, every 0 days at 0 hours)
(Auto, every 0 days at 0 hours).
Part I: Control Unit
Part I: 4-1