Teledyne 3010PA Operating Instructions Manual page 68

Percent oxygen analyzer
Table of Contents

Advertisement

Oxygen Analyzers
Span In
Zero In
Sample In
In vacuum service the
restrictor should be
placed here.
Exhaust Out
Restrictor
The Model 3010P sample system is designed and built to ensure that
the oxygen concentration of the gas is not altered as it travels through the
sample system. The sample encounters almost no dead space. This mini-
mizes residual gas pockets that can interfere with analysis at low oxygen
levels.
The metal restrictor upstream from the cell helps manage the flow
through the system (1 SLPM at 20 psig for nitrogen). It is corrosion resistant.
The sample or calibration gas flowing through the system is monitored
by a flowmeter downstream from the cell. The sample system for the stan-
dard instruments incorporates
connections on the Gas Control Panel. For metric system installations, 6 mm
adapters are supplied.
For -Vacuum Service, the restrictor is located downstream of the
flowmeter. The restrictor is installed in the exhaust poert on the gas panel.
Solenoid
Valves
In normal service the
restrictor should be
placed here.
Figure 2-4: Flow Diagram
inch tube fittings for sample inlet and outlet
1/4
Part II: Analysis Units
Components in the shaded area are in
the -C option (internal control valves)
only and are not shown in the piping
diagram above.
Cell
Flowmeter
Part II: 2-7