Page 1
Trace Oxygen Analyzer OPERATING INSTRUCTIONS FOR MODEL 3000TA-XL-EU Trace Oxygen Analyzer P/N M69603 9/24/10 DANGER Toxic gases and or flammable liquids may be present in this monitoring system. Personal protective equipment may be required when servicing this instrument. Hazardous voltages exist on certain components internally which may persist for a time even after the power is turned off and disconnected.
Page 2
Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
Page 4
19" panel and ready to mount in a standard rack. 3000TA-XL-I Web enabled version. Sensor Options Available for the Instrument with the Above Serial Number: Insta-Trace B2CXL (Default) L2CXL Teledyne Analytical Instruments...
Page 5
Trace Oxygen Analyzer Important Notice Model 3000TA-XL-EU complies with all of the requirements of the Commonwealth of Europe (CE) for Radio Frequency Interference, Electromagnetic Interference (RFI/EMI), and Low Voltage Directive (LVD). The following International Symbols are used throughout the Instruction Manual. These symbols are visual indicators of important and immediate warnings and when you must exercise CAUTION while operating the instrument.
Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only. NOTE: Additional information and comments regarding a Symbol specific component or procedure are highlighted in the form of a note. Teledyne Analytical Instruments...
Page 7
Manuals do get lost. Additional manuals can be obtained from Teledyne at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com.
Page 8
Since the use of this instrument is beyond the control of Teledyne, no responsibility by Teledyne, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
2.3 Sample System 2.4 Electronics and Signal Processing Installation ................... 19 3.1 Unpacking the Analyzer 3.2 Mounting the Analyzer 3.3 Rear Panel Connections 3.3.1 Gas Connections 3.3.2 Electrical Connections 3.3.2.1 Primary Input Power 3.3.2.2 50-Pin Equipment Interface Connector Teledyne Analytical Instruments...
Page 10
4.3.3.2 Installing or Changing the Password 4.3.4 Logout 4.3.5 System Self-Diagnostic Test 4.3.6 Version Screen 4.4 Calibration of the Analyzer 4.4.1 Zero Cal 4.4.1.1 Auto Mode Zeroing 4.4.1.2 Manual Mode Zeroing 4.4.1.3 Cell Failure 4.4.2 Span Cal Teledyne Analytical Instruments...
Page 11
6.2.1 Storing and Handling Replacement Cells 6.2.2 When to Replace a Cell 6.2.3 Removing the Micro-Fuel Cell 6.2.4 Installing a New Micro-Fuel Cell 6.2.5 Cell Warranty 6.3 Fuse Replacement 6.4 System Self Diagnostic Test 6.5 Major Internal Components Teledyne Analytical Instruments...
Page 12
3000TA-XL-EU 6.6 Cleaning 6.7 Troubleshooting Appendix ..................83 A-1 Specifications A-2 Recommended 2-Year Spare Parts List A-3 Drawing List A-4 19-inch Relay Rack Panel Mount A.5 Application notes Material Safety Data Sheet Teledyne Analytical Instruments...
Page 13
Figure 2-5B: Flow Diagram—Low Pressure-Model 3000TA-XL-VS14 Figure 2-5C: Flow Diagram—Sample Under Pressure-Model 3000TA-XL-CV Figure 2-6: Power supply and PCB Location Figure 2-7: Block Diagram of the Model 3000TA-XL-EU Electronics Figure 3-1: Front Panel of the Model 3000TA-XL-EU Figure 3-2: Required Front Door Clearance...
Page 14
Figure 5-7: Self-Test Window Figure 6-1: Removing the Micro-Fuel Figure 6-2: Installing the Insta-Trace Cell Figure 6-3: Removing Fuse Block from Housing Figure 6-4: Installing Fuses Figure 6-5: Rear Panel Removal Figure A-1: Single and Dual 19" Rack Mounts Teledyne Analytical Instruments...
Page 15
Table 3-3: Remote Calibration Connections Table 3-4: Range ID Relay Connections Table 3-5: Commands via RS-232 Input Table 3-6: Required RS-232 Options Table 4-1: Output Signals Table 4-2: Range ID Output Voltage Table 6-1: Self Test Failure Codes Teledyne Analytical Instruments...
Oxygen Analyzer is a versatile microprocessor-based instrument for detecting oxygen at the parts-per-million (ppm) level in a variety of gases. This manual covers the Model 3000TA-XL-EU General Purpose flush-panel and/or rack-mount units only. These units are for indoor use in a nonhazardous environment.
Convenient and versatile, steel, flush-panel or rack- mountable case with slide-out electronics drawer. 1.4 Model Designations 3000TA-XL: Standard model for sample under pressure 3000TA-XL-VS: Instrument configured for Vacuum Service 3000TA-XL-CV: Instrument with calibration valves 3000TA-XL-I: Web enabled version Teledyne Analytical Instruments...
Introduction 1.5 Front Panel (Operator Interface) The standard 3000TA-XL-EU is housed in a rugged metal case with all controls and displays accessible from the front panel. See Figure 1-1. The front panel has thirteen buttons for operating the analyzer, a digital meter, an alphanumeric display, and a window for viewing the sample flowmeter.
Page 20
Access Door: For access to the Micro-Fuel Cell, the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in with a narrow gauge tool. Accessing the main circuit board Teledyne Analytical Instruments...
2 concentration alarms and 1 system alarm. RS-232 Port Serial digital concentration signal output and control input. Remote Probe Used in the 3000TA-XL-EU for controlling external solenoid valves only. Remote Span/Zero Digital inputs allow external control of analyzer calibration.
Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 3000TA-XL is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day. Figure 1-2: Model 3000TA-XL Rear Panel Teledyne Analytical Instruments...
Micro-Fuel Cell: In the battery, all reactants are stored within the cell, whereas in the Micro-Fuel Cell, one of the reactants (oxygen) comes from outside the device as a constituent of the sample gas being analyzed. The Micro-Fuel Cell is therefore a hybrid Teledyne Analytical Instruments...
4 cm . The cathode has many perforations to ensure sufficient wetting of the upper surface with electrolyte, and it is plated with an inert metal. Teledyne Analytical Instruments...
HALF REACTION: – – → 4OH + 2H O + 4e (cathode) (Four electrons combine with one oxygen molecule—in the presence of water from the electrolyte—to produce four hydroxyl ions.) Teledyne Analytical Instruments...
Micro-fuel Cell input is constant, a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode, and this rate is Teledyne Analytical Instruments...
(less than ± 0.2 ppm oxygen). Depending upon the application, zeroing may still be used to compensate for the combined zero offsets of the cell and the electronics. Figure 2-3: Characteristic Input/Output Curve for a Micro-Fuel Cell Teledyne Analytical Instruments...
Depending on the mode of operation either sample or calibration gas is delivered. The Model 3000TA-XL-EU sample system is designed and fabricated to ensure that the oxygen concentration of the gas is not altered as it travels through the sample system.
Trace Oxygen Analyzer Operational Theory 2.4 Electronics and Signal Processing The Model 3000TA-XL-EU Trace Oxygen Analyzer uses an 8031 microcontroller with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the analyzer.
Operational Theory 3000TA-XL-EU Figure 2-7: Block Diagram of the Model 3000TA-XL-EU Electronics Teledyne Analytical Instruments...
Page 33
4-20 mA DC and the 0-1 V DC analog concentration signal outputs, and the analog range ID outputs. Signals from the power supply are also monitored, and through the microprocessor, the system failure alarm is activated if a malfunction is detected. Teledyne Analytical Instruments...
Page 34
Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments...
Immediately report any damage to the shipping agent. 3.2 Mounting the Analyzer The Model 3000TA-XL-EU is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. The standard model is designed for flush panel mounting. Figure 3-1 is an illustration of the 3000TA-XL-EU standard front panel and mounting bezel.
Installation 3000TA-XL-EU Figure 3-1: Front Panel of the Model 3000TA-XL-EU All operator controls, except the flow control needle valve, are mounted on the control panel, which is hinged on the left edge and doubles as the door that provides access to the sensor and cell block inside the instrument.
Trace Oxygen Analyzer Installation 3.3 Rear Panel Connections Figure 3-3 shows the Model 3000TA-XL-EU rear panel. There are ports for gas inlet and outlet, power, communication, and both digital and analog concentration output via the 50-pin Equipment Interface Connector. Figure 3-3: Rear Panel of the Model 3000TA-XL-EU 3.3.1 Gas Connections...
PROVIDE EXCELLENT EMI/RFI PROTECTION. THE PLUG CASE MUST BE CONNECTED TO THE CABLE SHIELD, AND IT MUST BE TIGHTLY FASTENED TO THE ANALYZER WITH ITS FASTENING SCREWS. ULTIMATELY, IT IS THE INSTALLER WHO ENSURES THAT THE CONNECTIONS PROVIDE ADEQUATE EMI/RFI SIELDING. Teledyne Analytical Instruments...
Figure 3-4: Equipment Interface Connector Pin Arrangement 3.3.2.2.1 Analog Outputs There are four DC output signal pins—two pins per output. For polarity, see Table 3-1. The outputs are: Teledyne Analytical Instruments...
Each relay has both normally open and normally closed contact connections. The contact connections are shown in Table 3-2. They are capable of switching up to 3 amperes at 250 V ac into a resistive load. The connectors are: Teledyne Analytical Instruments...
3.3.2.2.4 Cal Contact This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.) Table 3-3: Remote Calibration Connections Function + Remote Zero – Remote Zero + Remote Span – Remote Span Teledyne Analytical Instruments...
Cal Contact 3.3.2.2.5 Remote Calibration Protocol To properly time the Digital Remote Cal Inputs to the Model 3000TA-XL-EU Analyzer, the customer's controller must monitor the Cal Relay Contact. When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.
Pins 13 (+) and 29 (–). 3.3.2.2.8 Remote Valve Connections The 3000TA-XL-EU is a single-chassis instrument, which has no Remote Valve Unit. Instead, the Remote Valve connections are used as a method for directly controlling external sample/zero/span gas valves.
In addition, each individual line has a series FET with a nominal ON resistance of 5 ohms (9 ohms worst case). This can limit the obtainable voltage, depending on the load impedance applied. See Figure 3-6. Figure 3-6: FET Series Resistance Teledyne Analytical Instruments...
RS-232, until st<enter> is sent again. The RS-232 protocol allows some flexibility in its implementation. Table 3-6 lists certain RS-232 values that are required by the 3000TA- XL-EU implementation. Table 3-6: Required RS-232 Options Parameter Setting Baud 2400 Teledyne Analytical Instruments...
B-2CXL Micro-fuel Cell or as an option, Teledyne‘s Insta- Trace cell. An existing instrument can also be retrofitted to use the Insta- Trace cell using Teledyne retrofit kit P/N C-71792-A. This kit contains a cell holder and an Insta-Trace cell.
This cell is capable of measuring below 1 ppm of oxygen within 15 minutes of installation. By reducing the recovery time from several hours to just minutes, users will no longer have to divert unanalyzed product or switch analyzers. Teledyne Analytical Instruments...
Page 49
8. Push the gate on the cell block down so that the slots on the side of the gate engage the locating screws on the side of the block. This forces the holder into position and forms a gas- tight seal. Teledyne Analytical Instruments...
Before plugging the instrument into the power source: Check the integrity and accuracy of the gas connections. Make sure there are no leaks. Check the integrity and accuracy of the electrical connections. Make sure there are no exposed conductors. Teledyne Analytical Instruments...
Page 51
Trace Oxygen Analyzer Installation Check that inlet sample pressure is within the accepted range (see Section 3.3.1). Power up the system, and test it by repeating the Self- Diagnostic Test as described in Chapter 4, Section 4.3.5. Teledyne Analytical Instruments...
Set alarm setpoints, and modes of alarm operation (latching, failsafe, etc). Before you configure your 3000TA-XL-EU, these default values are in effect: Ranges: LO = 1 ppm, MED = 10 ppm, HI = 100 ppm.
HI or LO acting, latching, and/or failsafe. Range. Used to set up three analysis ranges that can be switched automatically with auto-ranging or used as individual fixed ranges. Teledyne Analytical Instruments...
Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at the appropriate point in the procedure, in a Arial Narrow Bold type style. Pushbutton names are printed in Oblique type. Teledyne Analytical Instruments...
Show Negative: The operator selects whether the display can show negative oxygen readings or not. TRAK/HLD: The operator sets whether the instrument analog outputs track the concentration change during calibration and sets a time delay for the concentration alarms after calibration. Teledyne Analytical Instruments...
HOLD TRACK HOLD calibration regardless of selecting . But when selected, the concentration alarms will remain frozen for the time TRAK/HLD displayed in the second line of the menu after the analyzer returns to the Analyze mode. Teledyne Analytical Instruments...
Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control where possible. The internal clock in the Model 3000TA-XL-EU is acurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day. To setup an Auto–Cal cycle:...
Enter the old password first. If the default password is in effect, pressing the Enter button will enter the default TETAI password for you. Press System to enter the System mode. TRAK/HLD Auto—Cal PSWD Logout More Teledyne Analytical Instruments...
NSTALLING OR HANGING THE ASSWORD If you want to install a password, or change an existing password, proceed as above in Entering the Password. When you are given the opportunity to change the password: Change Password? <ENT>=Yes <ESC>=No Teledyne Analytical Instruments...
Page 61
& < > When you have finished typing the new password, press Enter. A verification screen appears. The screen will prompt you to retype your password for verification. A A A A A Retype PWD To Verify Teledyne Analytical Instruments...
To log out, press the System button to enter the System function. TRAK/HLD Auto—Cal PSWD Logout More Use the ◄► arrow keys to position the blinking over the Logout function, and press Enter to Log out. The screen will display the message: Protected Until Password Reentered Teledyne Analytical Instruments...
Trace Oxygen Analyzer Operation 4.3.5 System Self-Diagnostic Test The Model 3000TA-XL-EU has a built-in self-diagnostic testing routine. Pre-programmed signals are sent through the power supply, output board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 3.
3 to 5 minutes. Note: The recovery time is significantly less, typically 30 minutes or less, when the Insta-Trace sensor is used. Refer to Section 3.4 for information on retrofitting your system to use an Insta-Trace sensor. Teledyne Analytical Instruments...
<ENT> To Begin Press Enter to begin zeroing. #### PPM Zero Slope=#### ppm/s The beginning zero level is shown in the upper left corner of the display. As the zero reading settles, the screen displays and updates Teledyne Analytical Instruments...
30 seconds. When Slope is close enough to zero, press Enter. In a few seconds, the screen will update. Once zero settling is completed, the information is stored in the microprocessor, and the instrument automatically returns to the Analyze mode. Teledyne Analytical Instruments...
Cell failure in the 3000TA-XL-EU is usually associated with inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the instrument will still Zero, but the 3000TA-XL-EU system alarm trips, and the LCD displays a failure message.
Cal hold: 5 min This menu allows the operator to set the time the analyzer should be held in the AUTO span mode. It does not have any effect in the MANual mode. Just press Enter key to continue. Teledyne Analytical Instruments...
To reset the alarm and the flag message, the unit must be turned off by cycling the standby key . It will reset if the next span cycle is correct. Teledyne Analytical Instruments...
Consider this before replacing the cell. 4.5 Switching of Sample Streams The Model 3000TA-XL-EU may be used to monitor more than one type of sample gas, such as nitrogen and helium, or two streams of nitrogen containing low (less than 1 ppm) and high (5000 to 10,000 ppm) concentrations of oxygen.
Consult Teledyne for further details. 4.6 The Alarms Function The Model 3000TA-XL-EU is equipped with 2 fully adjustable concentration alarms and a system failure alarm. Each alarm has a relay with a set of form ―C" contacts rated for 3 amperes resistive load at 250 VAC.
Page 72
Dft—N Fs—N Ltch—N Five parameters can be changed on this screen: AL–1 #### Value of the alarm setpoint, ppm (oxygen); this value can be set from 0.00 to 10,000.00 ppm Out-of-range direction, Dft–Y/N Defeated? (Yes/No) Teledyne Analytical Instruments...
Low = 0–1 ppm Med = 0–10 ppm High = 0–100 ppm. The Model 3000TA-XL-EU is set at the factory to default to autoranging. In this mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity.
The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range. To switch from autoranging to fixed range analysis, enter the range function by pressing the Range button on the front panel. Teledyne Analytical Instruments...
Analyze button at any time to return to analyzing your sample. 4.9 Signal Output The standard Model 3000TA-XL-EU Trace Oxygen Analyzer is equipped with two 0–1 VDC analog output terminals accessible on the back panel (one concentration and one range ID), and two isolated 4–20 mA DC current outputs (one concentration and one range ID).
To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current when using the current outputs) to represent a particular range. The following table gives the range ID output for each analysis range: Teledyne Analytical Instruments...
Micro-fuel Cell (see Section 3.4) 4.10 The IP Address Function This function is only available on web enabled instruments (3000TA-XL-EU-I) and is used to set the IP address of the embedded Ethernet web server coprocessor. To set the IP address, enter the system menu:...
Page 78
The 3000-I series analyzer microcontroller does not query the Ethernet Web Enabled co-processor on its current IP address. In other words, The IP address displayed is not necessarily the IP address of the Ethernet Web Enabled co-processor, Consider this if when replacing the Ethernet PCB. Teledyne Analytical Instruments...
The following three steps will web enable your instrument: 1. Follow the instructions in Chapter 4 Section 4.10 to set the IP address from the front panel of the 3000TA-XL-EU-I Trace Oxygen Analyzer. 2. Connect the analyzer to a network through the RJ45 connector on the side of the instrument.
On the top of the web page there are six functions that can be accessed by clicking on the six buttons. They are: Span Zero Alarms Ranges Teledyne Analytical Instruments...
IP address. 5.2 Remote Calibration The 3000TA-XL-EU-I can be zeroed and spanned remotely, but to do so, the instrument must have calibration valves under its control. Refer to Section 3.3.2.2.8 for remote probe connector outputs that can be used to control calibration valves.
'mode' will switch from 'analyzing' to 'getting zero'. Check the message window to find out when calibration is finished. If manual calibration was chosen, the operator must click the Finish button to force the instrument calibration. Figure 5-3: Zero Calibration Window Teledyne Analytical Instruments...
It takes about five seconds for the update to occur. Do not attempt adjustments while ‗Getting Data‘ is displayed in the message window. Repeat the process for adjusting the second alarm. Figure 5-4: Set Alarms Window Teledyne Analytical Instruments...
Web Implementation 5.4 Set the Ranges Remotely With the 3000TA-XL-EU-I, it is possible to change range settings remotely. Click on the Range button and the range menu will appear on the left side of the web page. After a few seconds, the data area will show the current settings of the instrument.
By clicking the System button some settings found in the system menu from the front of the control panel can be set, such as: Zero Offset, Alarm Delay, and Auto Calibration. Description of these settings can be found in Chapter 4. Figure 5-6: System Window Teledyne Analytical Instruments...
'self-test'. The self-test should not take more than one minute for a 3000TA-XL-EU-I. During this time, do not click on any buttons. At the end of the test, the results will be displayed on the left side of the web page.
Page 88
Web Implementation 3000TA-XL-EU Teledyne Analytical Instruments...
6.2.1 Storing and Handling Replacement Cells To have a replacement cell available when it is needed, TAI recommends that one spare cell be purchased 9-10 months after commissioning the 3000TA-XL-EU, or shortly before the end of the cell's one year warranty period. CAUTION: DO NOT STOCKPILE CELLS.
REACH ZERO AFTER INSTALLATION (1-2 WEEKS!). 6.2.2 When to Replace a Cell Cell failure in the 3000TA-XL-EU is usually characterized by a significant increase in the zero offset of more than 1.0 ppm oxygen or the inability to calibrate the sensor. When this occurs, the 3000TA-XL- EU will still zero but the system alarm trips, and the LCD displays a failure message.
3. Push the gate on the cell block down so that the slots on the side of the gate engage the locating screws on the side of the block. This forces the holder into position and forms a gas- tight seal. Teledyne Analytical Instruments...
6.2.5 Cell Warranty The Class B-2CXL and the Insta-Trace Micro-fuel Cells are used in the Model 3000TA-XL-EU. These cells are warranted for 6 months from the date of shipment. Note any Addenda attached to the front of this manual for special information applying to your instrument.
1. Place small screwdriver in notch, and pry cover off, as shown in Figure 6-3. Figure 6-3: Removing Fuse Block from Housing 2. To change between American and European fuses, remove the single retaining screw, flip Fuse Block over 180 degrees, and replace screw. Teledyne Analytical Instruments...
3. Use the ◄► arrow keys to move to Self-Test, and press Enter. The following failure codes apply: Table 6-1: Self Test Failure Codes Power 5 V Failure 15 V Failure Both Failed Analog DAC A (0–1 V Concentration) DAC B (0–1 V Range ID) Both Failed Teledyne Analytical Instruments...
The gas piping is illustrated in Figures 2-4 and 2-5 and the major electronic components locations are shown in Figure 2-6, in Chapter 2. WARNING: S ee warnings on the title page of this manual Figure 6-5: Rear Panel Removal (Remove only those screws marked with an X) Teledyne Analytical Instruments...
DO NOT wipe front panel while the instrument is controlling your process. 6.7 Troubleshooting Erratic readings of the Oxygen concentration as reported by Problem: the analyzer. Possible Cause: The analyzer may have been calibrated in an inaccurate fashion. Teledyne Analytical Instruments...
Page 98
Turn the analyzer off, then back on again. Press the System key when prompted by the analyzer "Press System for Default Values". This will return the analyzer to its default settings in calibration and zero values. Now proceed to carefully calibrate and zero the analyzer. Teledyne Analytical Instruments...
• Flush panel mount (Standard) • Relay rack mount. Contains either one or two instruments in one 19" relay rack mountable plate (Optional). Sensor: Teledyne B-2CXL trace analysis Micro- fuel Cell. Insta-Trace Micro-fuel Cell (option) Cell Block: 316 stainless steel.
Page 100
Analog outputs: 0-1 VDC percent-of-range, 0-1 VDC range ID. 4-20 mA DC (isolated) percent-of-range, 4-20 mA DC (isolated) range ID. Dimensions: 19 cm high, 24.9 cm wide, 31 cm deep (6.96 in high, 8.7 in wide, 12.2 in deep). Teledyne Analytical Instruments...
(if available) and the model and serial number of the instrument for which the parts are intended. Orders should be sent to: TELEDYNE Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580 Phone (626) 934-1500, Fax (626) 961-2538 Web: www.teledyne-ai.com...
For operation at cell pressures other than atmospheric, care must be taken to avoid changing the sample pressure too rapidly or cell Teledyne Analytical Instruments...
Section I - Product Identification Product Name: Micro-fuel Cells Mini-Micro-fuel Cells Super Cell, all classes except T-5F Electrochemical Oxygen Sensors, all classes Manufacturer: Teledyne Electronic Technologies Analytical Instruments Address: 16380 Chestnut Street, City of Industry, CA 91749 Phone: (626) 961-9221...
Page 106
Carcinogenicity: NTP Annual Report on Carcinogens: Not listed LARC Monographs: Not listed OSHA: Not listed Other health hazards: Lead is listed as a chemical known to the State of California to cause birth defects or other reproductive harm. Teledyne Analytical Instruments...
Page 107
The information is believed to be correct but does not purport to be all inclusive and shall be used only as a guide. Teledyne Analytical Instruments shall not be held liable for any damage resulting from handling or from contact with the above product.
Need help?
Do you have a question about the 3000TA-XL-EU and is the answer not in the manual?
Questions and answers