The Range Function; Setting The Analog Output Ranges - Teledyne 3010PA Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

Oxygen Analyzer
4.6
The Range function allows the operator to program up to three concen-
tration ranges to correlate with the DC analog outputs. If no ranges are
defined by the user, the instrument defaults to:
The Model 3010PA is set at the factory to default to autoranging. In this
mode, the microprocessor automatically responds to concentration changes
by switching ranges for optimum readout sensitivity. If the current range
limits are exceeded, the instrument will automatically shift to the next higher
range. If the concentration falls to below 85% of full scale of the next lower
range, the instrument will switch to that range. A corresponding shift in the
DC percent-of-range output, and in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays
on a fixed range regardless of the oxygen concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc (20 mA at the current output).
However, the digital readout and the RS-232 output of the concentra-
tion are unaffected by the fixed range. They continue to read accurately with
full precision. See Front Panel description in chapter 1.
The automatic air calibration range is always 0-25 % and is not pro-
grammable.

4.6.1 Setting the Analog Output Ranges

To set the ranges, enter the range function mode by pressing the
Range button on the front panel.
Use the < > arrow keys to blink the range to be set: low (L), medium
(M), or high (H).
Go to Ltch– and then press either two times or
(Toggle it to N and back to Y.)

The Range Function

Range
Limits
Low
0–1 %
Med
0–5 %
High
0–10 %.