Manual Mode Spanning - Teledyne 2000A-EU Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
arrow keys to toggle between
appears, blinking, on the display.
Press Enter to move to the next screen.
Use the < > arrow keys to toggle between the span concentration value
and the units field (%/ppm). Use the
units, as necessary. When you have set the concentration of the span gas you are
using, press Enter to begin the Span calibration.
The beginning span value is shown in the upper left corner of the display.
As the span reading settles, the screen displays and updates information on
. Spanning automatically ends when the span output corresponds, within
tolerance, to the value of the span gas concentration. Then the instrument auto-
matically returns to the analyze mode.
Press Span to start the Span function. The screen that appears allows
you to select whether the span calibration is to be performed automatically or
manually.
Use the
when
screen.
Use the < > arrow keys to toggle between the span concentration value
and the units field (%/ppm). Use the
units, as necessary. When you have set the concentration of the span gas you are
using, press Enter to begin the Span calibration.
Press Enter to enter the span value into the system and begin the span
calibration.
4.4.2.2

Manual Mode Spanning

keys to toggle between
appears, blinking, on the display. Press Enter to move to the next
Teledyne Analytical Instruments
and
span settling. Stop when
arrow keys change the value and/or the
and
span settling. Stop
arrow keys change the value and/or the
Operation 4
4-15